Problem
Probable Cause
Action
Interference in adjacent
devices such as image
artifacts on screen
Unknown
First make sure the problem is
related to the system by examine
the interference after turning the
system off.
If the interference is still observed:
▪
Re-orient or relocate the other
device(s).
▪
Increase the separation
between the equipment.
▪
Connect the system into an
outlet on a circuit different
from that to which the other
device(s) are connected.
If the problem persists, contact
IceCure™
Medical service.
“Long wash” popup
message
System pipes blockage
The user will have to wait until the
wash ends.
In order to have better indication in
regards to the end of the long wash
procedure, it’s better not to click
“O.K” when the message pops up,
but to wait until the message
disappear by itself. In case you
clicked “O.K” the screen freeze till
the end of the “long wash”
procedure.
Prepare for Treatment
button in the Main
Menu screen is disabled
•
Maintenance is needed
•
Washing is in progress
1)
Contact
IceCure™
Medical
service representative.
2)
Wait until the button becomes
enabled.