5.2.8.5
Manual Freeze mode
The manual freeze mode allows you to perform a Cryoablation procedure without setting
its duration or the number of
Freeze
-
Thaw
cycles in advance. You can choose manual
freeze mode by pressing the Manual Mode icon at the bottom of the Treatment Selection
screen.
Figure 79: Choosing the Manual Mode option
Operational stages
Safe Operation in Percutaneous Procedures
When performing percutaneous cryotherapy procedures, meticulous observation of the
incision site and the skin overlying the treatment site is required. Skin protection
techniques must be implemented to avoid thermal injury to the skin.
Skin protection techniques include, but are not limited to: dripping room-temperature
sterile saline on these areas; continuous monitoring of the growing ice ball by ultrasound;
utilizing an external thermocouple to measure skin temperature; injecting sterile saline or
local anesthetic between the skin and the ice ball formation; and placing moist gauze
between the skin and the cryoprobe.
Preliminaries
Before operating
ProSense™
cryoablation system, make sure you have completed all pre-
operational stages.
Warning
Insertion of the introducer or cryoprobe into the target tissue is performed
under the guidance of an appropriate imaging device and by an authorized
practitioner trained by
IceCure™
Medical.