IceCure™ Medical Ltd.
DMS-7063 rev. C
ProSense™
Confidential
User Manual
102
US User Manual
Figure 105: Replacement was completed after the Thaw ended
5.3.7.1
Replace the Dewar during Freeze
In case of low level of liquid nitrogen,
the system can’t continue to Freeze.
It will ask to
refill the Dewar in Freeze step (see fig. below)
Figure 106: Replacement during Freeze, in case of empty Dewar.
•
If you press
OK
, you can replace the Dewar and continue the procedure. The system will
move to Manual mode and wait in a
Thaw
step for instruction.
Pay attention that in Manual mode, you need to press the icon on the screen for each
Freeze
or
Thaw
step.
•
Replace Dewar as in paragraph 5.3.8.
•
If you press
Cancel
, you will get a message of "Low Liquid Nitrogen Level" (See a figure
below).