US User Manual
Warning
Removing the Dewar or placing it back within the system after refilling it must
only be done according to system instructions and with the carriage in the
bottom position. If the carriage is not in the bottom position, liquid nitrogen may
spill out.
•
Close the compartment door and press
Next
on the screen.
Figure 56: Correct insertion and positioning of the Dewar
–
bottom first
Make sure the Dewar positioned freely in the base.
Figure 57: Insertion of the Dewar
–
head first, leaning forward
Such insertion might cause wrong positioning of the Dewar, leaning forward.
Figure 58: Insertion of the Dewar
-
bottom first, leaning backwards
Wrong bottom first positioning of Dewar on the back of the elevator might cause Dewar leaning
backwards.