The hose can be safely put on the Holder arm once positioning of the Cryoprobe in the
tissue is completed.
The Holder base is designed to be positioned under the patient body while the holder arm
is adjustable at the side of the table.
The Holder is designed to adjust to a CT movable table.
It is the consideration of the user on which side and where to position the holder arm
according to the room and table conditions, except under the CT machine gantry.
Figure 33: Holder and its position within the CT machine (not available in some regions, e.g. China)
Foot pedal (Not available in some regions, e.g. China)
Pushing the pneumatic foot pedal is identical to activating/deactivating the
Freeze
step
and confirming actions of the active buttons on the screen:
Figure 34: Action button as displayed on screen (several of the options)
Figure 35: Foot pedal
Temperature sensor (TS)
The temperature sensor is a single use accessory that provides real time temperature
measurements when inserted in tissue. Use of temperature sensors is optional; the
decision to use a temperature sensor is according to clinical judgment.
Prior to starting a procedure, the temperature sensor is connected to the system and
inserted into the desired location within the tissue. Insert the sensor at location that gives
useful information about the progress of the freezing procedure. For example, the
temperature sensor can be placed near the boundary of a tumor to ensure that the entire
tumor has been frozen. Alternatively, the temperature sensor may be placed near an
organ that should not be frozen during the procedure to ensure that the ice front has not
advanced to this location.