49
US User Manual
Figure 37: Temperature display of sensor in the bottom-right corner of the screen
Warning
Temperature sensor, introducer, or cryoprobe insertion and navigation within
a tissue must be done under guidance of an appropriate imaging device.
Dewar storage cases
The Dewar storage shelves are located at the back of the chassis, providing storage for
the Dewars. When placing the Dewars inside the storage shelf, make sure they are well
installed (see fig. below)
Figure 38: Dewar storage cases
–
the correct way to put the Dewars inside the storage shelf.
Safe transportation
When moving the system from one location to another the Dewars must be empty.
Operational details
This section describes the operational actions of the
ProSense™
cryoablation system that
support the functional operations described in Chapter 5.
•
Order liquid nitrogen in advance based on projected case load.