![IceCure ProSense DMS-7063 Скачать руководство пользователя страница 115](http://html1.mh-extra.com/html/icecure/prosense-dms-7063/prosense-dms-7063_user-manual_3378524115.webp)
IceCure™ Medical Ltd.
DMS-7063 rev. C
ProSense™
Confidential
User Manual
115
US User Manual
SYSTEM MAINTENANCE
General cleaning
Warning
Do not allow any liquid or humidity to enter the cryohandle. Always keep the
cover on the cryohandle.
Following each cryosurgical procedure, discard the single use devices (single-use
cryoprobe, single-use temperature sensor, cryohandle cover, and sleeves for the hose
and touch screen). Carefully clean the hose and the cryohandle according to the
cleaning instructions acceptable for permanent equipment in your institution.
All single use devices are considered to be medical waste and must be disposed of in
accordance with medical waste laws and hospital standards. Sharp objects such as the
cryoprobe, introducer and temperature sensor must be disposed of in an adapted
container.
Following a cryosurgical procedure, it is recommended to wipe the pedal with a damp
cloth and WARM soapy water to remove visible soil. Dry it with a clean cloth or with
forced air and wrap it in another clean cloth until disinfecting it according to the
disinfection acceptable standards in your institution.
If Holder was used during procedure, it is recommended to wipe the holder gripper
arm and base with gauze pads soaked with 70% alcohol to verify no remnants blood
or tissue, according to the disinfection acceptable standards in your institution.
Check the main unit for any remaining blood or tissue. In case these remnants are
detected, wipe the infected area with gauze pads soaked with 70% alcohol. Verify that
no visible remnants remain. Wipe the initially cleaned area with approved medical
professional surface disinfecting wipes and allow it to air dry. Please note that using a
high concentration of chlorine solution might damage the paint.
Thoroughly clean the
ProSense™
cryoablation system unit with a damp cloth and
warm soapy water. Dry it with a clean cloth. Clean the monitor screen with a soft
cloth and a mild window cleaning solution. Take special care to avoid spilling liquid
onto the system.
At the end of the day, carefully dry all items and store the system in an appropriate
place.