IACER S.r.l.
32 of 36 MNPG148 Rev. 02 02/07/15
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is inteded for use in the electromagnetic environment specified below. The customer or
the user of the device should assure that is used in suche environment
Immunity test
Test level EN
60601-1-2
Compliance level Electromagnetic
environment
–
guidance
Conducted RF
EN 61000-4-6
3 Vrms 150kHz
to 80MHz
3 Vrms 150kHz to
80MHz
Portable
and
mobile
RF
communications equipment should be
used no closer to any part of the device,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance:
d = 1,2
P 150kHz to 80MHz
d = 1,2
P 80 MHz to 800 MHz
d = 2,3
P 800 MHz to 2,5 GHz
where (P) is the maximum output
power rating of the transmitter in Watts
(W) according to the transmitter
manufacturer
and
(d)
is
the
recommended separation distance in
metres (m).
RF Radiata
EN 61000-4-3
3 Vrms 80MHz
to 2,5GHz
3 Vrms 80MHz to
2,5GHz
Field strangths from fixed RF transmitters, are determined by an electromagnetic site survey, should
be less than the complicance level in each frequency rage.
Interference may occur in the vicinity of equipment marked with the following symbol:
Recommended separation distances between portable and mobile communications
equipment and the device
The device is intended for the use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent electromagnetic
interferences by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the device as recommended below, according to the maximum output
power of the communication equipment.
Rated
maximum
power
of
the
transmitter (W)
Separation distance according to the frequency of the transmitter (m)
150kHz to 80MHz
d = 1,2
P
80MHz to 800MHz
d = 1,2
P
800MHz to 2GHz
d = 2,3
P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in Watts (W) according to
the transmitter manufacturer.
Note:
(1) At 80 MHz and 800 MHz the separation distance for the higher frequency range applies.
(2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.