IACER S.r.l.
29 of 36 MNPG148 Rev. 02 02/07/15
Use only the original battery charger or in any case the battery charger supplied by the
fabricant/distributor. Not open or modify the battery charger.
Cleaning
Clean the equipment from the dust using a soft cloth.
Resistant stains can be removed using a sponge soaked in solution of water and alcohol.
Device not subjected to sterilization.
Carriage and storage
Carriage precautions
MIO-CARE PRO is a portable device, so it does not need any particular carriage precautions.
However we recommend to put away MIO-CARE PRO and its accessories in their own bag after
every treatment.
Storage precautions
MIO-CARE PRO is protected till following environmental conditions:
In operation
Temperature
from +5 to + 40 °C
Rel. humidity
from 30 to 75%
Pressure
from 700 to 1060 hPa
Inside of the packaging
Temperature
from –5 to +55 °C
Rel. humidity
from 10 to 90%
Pressure
from 700 to 1060 hPa
Disposal
The equipment is subjected to WEEE regulations (see the symbol
on the label) concerning
separate waste collection: when disposing this product, please use the designed areas for disposing
electronic waste or contact the manufacturer.
Troubleshooting
If it is used in accordance with the instructions of the user manual, MIO-CARE PRO does not need a
particular regular maintenance.
If you find any malfunctioning using MIO-CARE PRO, please follow these instructions:
MIO-CARE PRO does not turn on and/or the display does not light up.
Check the battery status
and replace it if it is necessary (make reference to chapter "Battery replacement"). If the problem persists
contact the manufacturer.
MIO-CARE PRO does not transmit electric impulses
. Check that the cable jacks have been
inserted in the electrodes and that the plastic protection has been removed from the electrode. Check
that the cables have been connected correctly (connector well inserted in the device). Check that the
cables and the electrodes are not damaged. If the problem persists contact the manufacturer.
MIO-CARE PRO transmits low intensity or intermittent impulses.
Check the cables and the
electrodes are in good condition and replace them if it is necessary. If the problem persists contact the
manufacturer.
MIO-CARE PRO switches off during the operation
. It is suggested to replace the battery and start a
new treatment. If the problem persists contact the manufacturer.
MIO-CARE PRO does not allow the intensity adjustment or not keep the adjusted value and
reset.
It is suggested to replace the battery and start a new treatment. If the problem persists contact the
manufacturer.