EMC Tables
Electromagnetic emission
Emission test
Compliance
Electromagnetic
environment
–
guidance
RF emissions
Cispr 11
Group 1
The device uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment
RF emissions
Cispr 11
Class B
The device is suitable for use in all
establishments,
including
domestic
establishments
and
those
directly
connected to the public low-voltage power
supply network that supplies buildings
used for domestic purposes.
Electromagnetic immunity
The device is inteded for use in the electromagnetic environment specified below. The customer or
the user of the device should assure that is used in suche environment
Immunity test
Test level EN
60601-1-2
Compliance level
Electromagnetic environment
– guidance
Electrostatic discharge
(ESD)
EN 61000-4-2
6kV contact
8kV air
6kV contact
8kV air
Floors sholud be wood, concrete
or ceramic tile. If floor are
covered with syntethic material,
the relative humidity should be
at least 30%
Mains
power
electromagnetic field
EN 61000-4-8
3 A/m
3 A/m
Mains power quality should be
at that of a typical commercial or
hospital environment