IACER S.r.l.
17 of 36 MNPG148 Rev. 02 02/07/15
Important: for all of these programs, stimulation intensity must be set between the threshold of
impulse perception and the moment in which the impulse starts to cause discomfort. With the
exception of the “periarthritis” program, the muscles surrounding the area to be treated must not
contract, they should only produce slight “vibrations”.
BEAUTY Programs
Prg.
PHASE 1
PHASE 2
PHASE 3
B1
Total time 4 min
frequency 6 Hz
impulse width 200
s
Total time 15 min: (3Hz-
200
s x 7sec 80%+ 1Hz
200
s x 3 sec 100% +
20Hz-200
s x 5 sec 80%) x
60 cycles
Total time 10 min: (3Hz-
200
s x 7sec 80%+ 1Hz
200
s x 3 sec 100% + 30Hz-
200
s x 5 sec 80%) x 40 cycles
B2
Total time 4 min
frequency 6 Hz
impulse width 300
s
Total time 15 min: (3Hz-
300
s x 7sec 80%+ 1Hz
300
s x 3 sec 100% +
20Hz-300
s x 5 sec 80%) x
60 cycles
Total time 10 min: (3Hz-
300
s x 7sec 80%+ 1Hz 3-
00
s x 3 sec 100% + 30Hz-
300
s x 5 sec 80%) x 40 cycles
B3
Total time 4 min
frequency 6 Hz
impulse width 200
s
Total time 15 min: (3Hz-
200
s x 7sec 80%+ 1Hz
200
s x 3 sec 100% +
40Hz-200
s x 5 sec 75%) x
60 cycles
Total time 10 min: (3Hz-
200
s x 7sec 80%+ 1Hz
200
s x 3 sec 100% + 50Hz-
200
s x 5 sec 75%) x 40 cycles
B4
Total time 4 min
frequency 6 Hz
impulse width 300
s
Total time 15 min: (3Hz-
300
s x 7sec 80%+ 1Hz
300
s x 3 sec 100% +
40Hz-300
s x 5 sec 75%) x
60 cycles
Total time 10 min: (3Hz-
300
s x 7sec 80%+ 1Hz
300
s x 3 sec 100% + 50Hz-
300
s x 5 sec 75%) x 40 cycles
B5
Total time 4 min
frequency 6 Hz
impulse width 200
s
Total time 10 min: (3Hz-
200
s x 7sec 80%+ 1Hz
200
s x 3 sec 100% +
60Hz-200
s x 5 sec 70%) x
40 cycles
Total time 5 min: (3Hz-200
s
x 7sec 80%+ 1Hz 200
s x 3
sec 100% + 70Hz-200
s x 5
sec 70%) x 20 cycles
B6
Total time 4 min
frequency 6 Hz
impulse width 300
s
Total time 10 min: (3Hz-
300
s x 7sec 80%+ 1Hz
300
s x 3 sec 100% +
60Hz-300
s x 5 sec 75%) x
40 cycles
Total time 5 min: (3Hz-300
s
x 7sec 80%+ 1Hz 300
s x 3
sec 100% + 70Hz-300
s x 5
sec 75%) x 20 cycles
B7
Total time 4 min
frequency 6 Hz
impulse width 250
s
Total time 5 min:
frequency 12 Hz
impulse width 250
s (90%)
Total time 5 min: (5Hz-250
s
x 5sec 90%+ 30Hz 250
s x 5
sec 90% ) x 30 cycles
B8
Total time 4 min
frequency 12 Hz
impulse width 100
s
Total time 10 min: (5Hz-
100
s x 10sec 90%+ 20Hz
100
s x 5 sec 90% ) x 40
cycles