MultiSystem 5060
Operating Instructions Manual
© Hydrotechnik GmbH • All rights reserved
Rev. 1.8/120307EN
Page 49 of 55
7.6
Firmware update using the USB Stick
Attention
Malfunctions and loss of data possible!
Please ask Hydrotechnik whether your instrument is suited for the
planned firmware update. Possibly it is not possible to update an
early version to the latest directly. Save all data from the SD card
before starting the update.
You can use the USB stick to execute firmware updates of the Multi-
System 5060 in a fast and easy way:
1. Copy the new firmware version on the USB stick. The file must be
located on the top file system level, not in a directory.
2. Switch on the MultiSystem 5060 and wait until the measured val-
ues are displayed.
3. Plug the USB stick into the USB connector.
4. Open the device menu.
5. Open the setup menu with
.
6. Highlight the function „USB stick“ and press
.
7. The message „Read directory“ will be displayed and after a short
time, the data contained on the stick are displayed.
8. Highlight the file „60_00.BIN“ and press
.
9. Confirm the update with
(OK).
10. The new firmware version will be loaded what takes about 3 min-
utes.
Attention
Malfunctions and loss of data possible!
Do not switch off the instrument and do not unplug the USB stick
during the new firmware version is loaded. Otherwise there could be
malfunctions and loss of data.
11. After loading the firmware, the message „Update Firmware“ will be
displayed.
12. Confirm the update with
(yes).
13. The update will be executed, then the instrument will be switched
off automatically.
14. Unplug the USB stick.
15. Switch the instrument on. The current firmware version number
and the message „Execute reset“ will be displayed.
16. Confirm the reset with
(yes).
17. Confirm the operation languages with
.
18. Wait until the initialization has been executed completely.
19. Format the SD card.
Now you can use your MultiSystem 5060 with the new firmware ver-
sion.