6.3.3
Configure trigger output (Ch10)
You can manage an external control dependant on incidants with the
trigger output. You have to define five parameters:
Variable
shows the internal measurand of the trigger channel
Name
you may assign an individual name
State
source of the trigger event; INACTIVE: trigger off,
CHANNEL: meas. channel is supervised for the trigger
event; MEM-TRIG: trigger is switched if triggering was
detected during storing [allows to xynchronise several
instruments: master: storing trigger event X (e.g. p1 >
200) – trigger output: SP_TRIG; slaves: storing trigger
event E1], MANUAL: trigger output is switched manu-
ally by pressing a key
Variable
select which measuring channel shall bring the trigger
event
Condition
for trigger input ON/OFF, for measuring channels
GREATER/LOWER
Value
for measuring channels, e.g. 200
6.3.4 Configure
special channels (Ch11 ... 24)
The special channels are used to combine the measured values of
several sensors mathematically and do calculations with it, or to be
configured as input channels for the CAN bus or the RS232 interface.
Calculation
choose between the different occupations of the chan-
nel (see further below)
Variable
is entered automatically when using pre-programmed
formulas and cannot be edited; for individual formulas
and occupation with CAN or RS232 you may define the
variable here that is provided on this channel
Index variable is manual indexing is set in the setup menu (see section
6.6.14 on page 31), you may enter the index number of
the channel here
Unit
is entered automatically when using pre-programmed
formulas and cannot be edited; define the unit for chan-
nels with individual formulas, CAN, or RS232 occupa-
tion
Name
you can enter an individual name for the channel
Align.Diff
this functions automatically determines the measured
value difference between the selected channels and
use it as offset
Formula
enter the desired formula here (only displayed if “Calcu-
lation” is set to FORMULA, see below)
Additional functions
loads stored channel parameters from the database
stores the current channel parameters in the database
saves the channel settings and leaves the submenu