MultiSystem 5060
Operating Instructions Manual
© Hydrotechnik GmbH • All rights reserved
Rev. 1.8/120307EN
Page 15 of 55
5.5
Enter sensor parameters
Note
If you have connected ISDS sensors, the sensor parameters will be
detected automatically when the instrument is switched on. Then you
can skip this section.
Note
If you have connected sensors without ISDS function, you will have to
program the sensor parameters manually. You find the required in-
formation e.g. on the type plate of the calibration protocol of your
sensor.
Open channel menu:
Highlight channel:
Start programming:
Highlight menu item:
Select menu item:
Highlight setting:
or enter value, e.g. 12.5:
Confirm setting or value:
Save changes:
Available measurands
The instrument is able to process ~ 40 different measurands like pressure, volume flow rate, temperature and ro-
tational speed. Assure to select the measurand and unit in accordance to the sensor.
Index variable
If several channels are programmed with the identic measurand, these will be indexed consecutively. The auto-
matic indexing can be disabled in the device menu to allow manual assignment of index numbers.
Name
You can assign an individual name to each channel.
Signal types
Select between „0/20 mA“ – „4/20 mA“ – „0/10 V“ – „± 10 V“ –
„0,5/4,5 V“ – „1/5 V“ – „2/10 V".
Measuring range
Enter the beginning and end of the measuring range and confirm with
.