Changing the filter membrane
Changing the filter membrane
Before you remove the filter membrane, the analysis liquid must have been
completely sucked away.
To remove the filter membrane, proceed as follows:
1. Hold the holder with one hand on
the lower part.
With your other hand, undo the
union nut counter-clockwise.
2. Put down the lower part together
with the filter membrane in its
holder.
3. Remove the filter membrane for
analysis and label it as shown in
the example on page
Fehler!
Textmarke nicht definiert.
.
CTU 1000 Series
en(us)
Page 55 / 92
BeWa CTU1000 3229578p en-us 2017-04-25.doc
2017-04-25