Safety Information
Proper/designated use
Claims for defects or liability, regardless of the legal foundation, do not apply
with incorrect or improper installation, commissioning, usage, handling,
storage, maintenance, repair, use of unsuitable components or other
circumstances for which HYDAC is not responsible.
HYDAC is not responsible for the installation, integration, selection of
interfaces to / into your system nor for the use or functionality of your system.
Use the unit only for the application described in the following.
The ContaminationTest Unit CTU is a test bench for analyzing the technical
cleanliness of components and systems in accordance with directive VDA 19
(ISO 16232 or 18413).
DANGER
Only with soluble
test liquid: "G 60 Spezial"
Flammable gas-air mixture in the analysis
chamber
Risk of burns
►
Use only test liquids with a flash point ≥
60°C
or with a lower explosion limit in excess of
0.6% by volume.
CTU 1000 Series
en(us)
Page 11 / 92
BeWa CTU1000 3229578p en-us 2017-04-25.doc
2017-04-25