HMG 3010
Page 52
Edition 2013-03-26 - V04 R01
Part No.: 669855
•
In order to also manually generate a
measured value line, set a second event
to
Actuation of button/key
.
•
Automatic measurement:
Deactivated
When measurement starts, a new menu
window
Event Log Recording — Status
opens, enabling you to track the measure-
ment on the basis of the settings. Data is
recorded only when an event occurs.
The sampling rate for the event log is auto-
matically determined by the HMG in keeping
with the number of active measurement
channels:
•
0.1 ms max. 2 analogue input channels
•
0.2 ms max. 4 analogue input channels
•
0.5 ms all 10 input channels
•
1.0 ms for SMART sensors
For the above example this means that the
stamping press is switched on, the pressure
reaches 131 bar and drops back to 0 bar.
When reaching 5 bar in the falling direction,
a measured value line is generated and the
Min/Max values are reset. The last Max val-
ue is retained in the measured value line. A
die closing pressure of 129 bar is reached in
the next cycle. This value is also written to a
new line as the pressure drops to 5 bar.
The function bar offers you the option of
having the current measured values dis-
played (
Measured Values
) or a
Table
shown.
The
Table
option
will list the
data records
of each individual event
line by line. Apart
from the date and time of the respective re-
cording, the trigger event and the Min and
Max values for the active sensors are
shown. The
most recent data record
is
shown in the
bottom line
of the table
.