HMG 3010
Page 35
Edition 2013-03-26 - V04 R01
Part No.: 669855
PC Connection
The
HMGWIN 3000
and
CMWIN
software
enables you to make online measurements
and settings in the HMG 3010, in addition to
further processing, archiving, documenting
and updating etc. Data transfer generally
takes place via the USB port.
You can also transfer data to the PC via the
HMG's serial port. To do this, select the
RS232(COM) setting along with the requisite
baud rate.
Ground potential of the sensor ports and
interfaces (USB/serial) are connected gal-
vanically. If sensors and interfaces are used
simultaneously
(online
measurement),
please ensure equipotential bonding to avoid
transient currents between the electricity
grids with different ground potential via
HMG, as the device could be destroyed.
Only for connection to CAN-bus:
CAN-Baud Rate [kBit]
Under the menu point
CAN-Baudrate [kBit]
you can specify the required Baud rate (from
10 to 1000 kBit) for the CAN-network (to
which the HMG 3010 is to be connected).
For all the setting options and operation for
CAN-Bus use, see Section 6.5‚ CAN Func-
tions.