HMG 3010
Page 31
Edition 2013-03-26 - V04 R01
Part No.: 669855
6.1.3 Zeroing Measurement Channels
You can re-zero the measurement channels
in the Settings menu. This function is used
to mask out backpressure buildups present
in the system. For example, if a pressure of
3 bar is present at a pressure transmitter
despite the fact that the system has been
switched off, this value can be used as the
new zero point for display purposes.
Re-zeroing can be done at any point
throughout the measurement range of a
measurement channel.
In the
Settings
menu, select
Zero meas-
urement channels
, and then confirm with
OK
. The screen on the right appears. You
are now prompted to select a measurement
channel. Select the channel (checkbox)
which you would like to zero using the arrow
keys, and then confirm with
OK
. Then select
Set
followed by
OK
— you have zeroed the
channel. You can delete the changed zero
point by pressing
Delete
.
All measurement channels which have
been re-zeroed are indicated by an arrow
in front of their designation for safety
reasons (see screenshot).
Caution!
A high pressure may still be present in a
measurement channel marked by an arrow
even though the display shows
0 bar
.
When removing a sensor or other system
components make sure the system has been
depressurized
first.
Failure to do so may result in serious in-
jury or death!