Field Application Department
2018 . 10 . 02 (r1)
25/63
more than 2 lines
OK
4) How to do calibration for IOP
[HNT-1/1P] Data issue_Solenoid Control
600(±10)
200(±10)
[Adjust the Solenoid Control parameter]
5. In case of the “
OK
”, perform the next procedure
- OK : Peak to Valley >= 2 lines
6. Set the pressure value using VR in the Main Board
- Minimum value : 200 -> VR8
- Maximum value : 600 -> VR9