[Check the Pachy calibration data]
13. After that, you can check whether the calibration is working properly or not
14. Press the Press the parameter you want to check
15. Press the left side “check”, you can see the CCT data at the bottom side
※ We will supply three Pachy model eyes (About 400, 600 and 800 um)
※ Tolerance of setup data should be under the ±5um
[HNT-1/1P] Data issue_Pachy Calibration
5) How to do check the Pachy calibration data