[Adjust the brightness for External LED]
※ The REF/KERATOMETER model eye should be used.
1. Check the LUSZ, LDSZ, RUSZ and RDSZ parameter and adjust this values if necessary
: LUSZ, LDSZ, RUSZ, RDSZ: Size (Brightness) -->
SIZE: 25~35 / Brightness: over 200
- Engineer Setup Mode → ③ Focus Led Setup → ③, ④ Adjust the External LED R/L
- Adjust the VR2(R) and 3(L) on the Mainboard using the plastic screwdriver (Clockwise: increase)
VR2(R)
5) How to adjust the LED brightness
[HNT-1/1P] Auto Tracking issue_EXT LED
VR3(L)