① [IOP Offset]
- Set the IOP offset value on the Measure Setup Mode
5
10
15
20
25
30
35
40
10
15
20
25
30
3) IOP Offset
[HNT-1/1P] Data issue_IOP Offset
※ If the measured value is uniformly high or low, you can
adjust the IOP Offset value to acquire the accurate data.
Ex) Measurement result: IOP10 -> 10 / IOP30 -> 28
IOP Offset -> Offset10: 10.0, Offset30: 32