Fit a new red lens into the mounting plate from the front. Install the LED into
the lens and push on a new retaining sleeve. Put new shrink sleeves onto the
wires from the alarm module and connect the wires to the indicator by means of
the pin/socket connectors. Slide the sleeves over the connectors and the LED
wires and apply hot air to shrink in position.
The alarm light LED spare part (No. BP384) is supplied as a complete
assembly which can be stripped down as described above to replace individual
components as desired.
5.3
Replacing the Audible Alarm
To remove the faulty audible alarm, pry it OFF. If this is difficult, leave it in
place and cut OFF the connections close to the body.
Attach the new alarm to the case in the correct position (or adjacent to the faulty
unit) using a cyanoacrylate adhesive.
Reconnect the electrical wires as shown in the wiring diagram using the method
described in paragraph 5.2.
5.4
Replacing the ON/OFF Switch
To remove the on/off switch, disconnect (pull OFF) the electrical connections,
squeeze together the retaining flanges located at the top and bottom of the
switch body and push the switch through the front panel.
To replace the switch, push in from the front ensuring the switch retaining
flanges snap into position.
Reconnect the electrical connections as per the wiring diagram.
F048
5-2
Issue 3
Mar 2003
Содержание Flowtron DVT
Страница 1: ...F048 Issue 3 Mar 2003 Huntleigh HEALTHCARE Flowtron DVT Service Manual ...
Страница 2: ...F048 Issue 3 Mar 2003 ...
Страница 6: ...Issue 3 iv F048 Mar 2003 ...
Страница 10: ...F048 1 4 Issue 3 Mar 2003 ...
Страница 12: ...F048 2 2 Issue 3 Mar 2003 ...
Страница 16: ...F048 3 4 Issue 3 Mar 2003 ...
Страница 18: ...F048 4 2 Issue 3 Mar 2003 ...
Страница 30: ...F048 5 12 Issue 3 Mar 2003 ...
Страница 32: ...F048 6 2 Issue 3 Mar 2003 ...
Страница 33: ...WIRING DIAGRAMS F048 1 1 Issue 3 Mar 2003 Flowtron DVT AC 500 ...
Страница 34: ...F048 1 2 Issue 3 Mar 2003 ...
Страница 37: ...F048 Issue 3 Mar 2003 ...