3.2.4
Checking the Pressure-Control Assembly
If the compressor is operating correctly and the correct pressure cannot be
obtained at the pump outlets, the rotor/stator assembly and the pressure-control
assembly must be checked:
1.
Remove the timer motor/gearbox assembly as described in Section 5.
2.
Compress the rotary-valve assembly and remove the drive pin.
3.
Check the bearing surfaces for cracks or bad scores. Also check the
security of the elbow connectors.
4.
Remove the pressure-control assembly as described in Section 5.
5.
Inspect the pressure control bags for leaks.
Any damaged parts must be replaced. It is recommended, in either case, that
the complete assembly is replaced.
Switch OFF the pump unit and disconnect the electric cord from the power
outlet before performing any repair work.
3.2.5
Loss of electrical power
Power loss can be due to the following:
1.
No electricity supply at wall outlet
2.
Faulty wall outlet
3.
Damaged plug or electric cord
4.
Defective fuse link
5.
Damaged internal wiring
Evaluate each potential fault and correct as necessary.
3.2.6
Technical assistance
For technical assistance, contact:
Service Department
Huntleigh Healthcare Inc.
40 Christopher Way
Eatontown,
New Jersey 07724
-3327
Telephone :
732 578 9898
Fax :
732 578 9889
F048
3-3
Issue 3
Mar 2003
Содержание Flowtron DVT
Страница 1: ...F048 Issue 3 Mar 2003 Huntleigh HEALTHCARE Flowtron DVT Service Manual ...
Страница 2: ...F048 Issue 3 Mar 2003 ...
Страница 6: ...Issue 3 iv F048 Mar 2003 ...
Страница 10: ...F048 1 4 Issue 3 Mar 2003 ...
Страница 12: ...F048 2 2 Issue 3 Mar 2003 ...
Страница 16: ...F048 3 4 Issue 3 Mar 2003 ...
Страница 18: ...F048 4 2 Issue 3 Mar 2003 ...
Страница 30: ...F048 5 12 Issue 3 Mar 2003 ...
Страница 32: ...F048 6 2 Issue 3 Mar 2003 ...
Страница 33: ...WIRING DIAGRAMS F048 1 1 Issue 3 Mar 2003 Flowtron DVT AC 500 ...
Страница 34: ...F048 1 2 Issue 3 Mar 2003 ...
Страница 37: ...F048 Issue 3 Mar 2003 ...