5.12 Replacing the Compressor Assembly Components
Unscrew and remove the 4 x M4 countersunk screws, remove the lid and gasket
and place the lid to one side. Discard the gasket and replace with new item.
5.12.1
Silencer Bag Assembly
Disconnect the silencer bag outlet tubes from the four-way connector,
disconnect silencer bag inlet tubes from the valve-body assemblies.
Replacement is the reverse of above procedure.
5.12.2
Bellows/Armature Blade Assemblies/Pivot Points
Unclip bellow from valve-body assembly and extract armature blade from pivot
point using long nosed pliers.
If necessary, use a sharp pointed tool to remove the pivot points from the holes
in the compressor case.
F048
5-7
Issue 3
Mar 2003