Replacement is the reverse of the above. When reconnecting tubes to the
assembly ensure that there are no kinks and tubes are pushed fully onto their
mounts. Be careful not to strain or damage wires when re-attaching the alarm
module.
5.9
Replacing the Electric Cord
The electric cord is fitted with a molded non-rewireable plug. This assembly
is not repairable and, if faulty, must be replaced.
Do not attempt to replace the plug with a rewireable type as this will negate
any warranty.
When an unserviceable power cord fitted with a non-rewireable plug is
removed from the equipment, the cord and plug must be destroyed and
disposed of safely. To avoid the danger of electric shock the plug must
never be inserted in any power outlet.
Remove the ground wire from the side of the compressor case and carefully lift
and lay aside the compressor assembly. Remove the two self-tapping screws
attaching the clamp plate to the bottom of the case. Disconnect the wires from
the terminal block, remove and dispose of the unserviceable electrical cord.
Thread the new cable through the grommet in the case and connect to the
terminal block as shown in the wiring diagram. Fit the clamp plate with two
self-tapping screws, ensuring that the wires are not strained.
Reposition the compressor assembly.
F048
5-5
Issue 3
Mar 2003
Содержание Flowtron DVT
Страница 1: ...F048 Issue 3 Mar 2003 Huntleigh HEALTHCARE Flowtron DVT Service Manual ...
Страница 2: ...F048 Issue 3 Mar 2003 ...
Страница 6: ...Issue 3 iv F048 Mar 2003 ...
Страница 10: ...F048 1 4 Issue 3 Mar 2003 ...
Страница 12: ...F048 2 2 Issue 3 Mar 2003 ...
Страница 16: ...F048 3 4 Issue 3 Mar 2003 ...
Страница 18: ...F048 4 2 Issue 3 Mar 2003 ...
Страница 30: ...F048 5 12 Issue 3 Mar 2003 ...
Страница 32: ...F048 6 2 Issue 3 Mar 2003 ...
Страница 33: ...WIRING DIAGRAMS F048 1 1 Issue 3 Mar 2003 Flowtron DVT AC 500 ...
Страница 34: ...F048 1 2 Issue 3 Mar 2003 ...
Страница 37: ...F048 Issue 3 Mar 2003 ...