PREVENTIVE MAINTENANCE
The following preventive maintenance should be performed every 12 months
4.1
Pump Unit
Replace the following components:
1.
Silencer bag
2.
Pivot points
3.
Bellow arms
4.
Valve bodies
5.
Compressor sponge filter
6.
Compressor gasket
Check:
a)
Base is free from debris
b)
Filter sponge position correct
c)
Compressor seating
d)
Electrical connections are secure and completely insulated
e)
Pressure control knob is tight
f)
Tubes are fully pushed on and are free from kinks
g)
Tubes and electrical insulation are not damaged
h)
Power switch illuminates
i)
Wiring is tidy
j)
Air is alternating
k)
Outlet tubes and snaplock connectors are secure and fitted
correctly
4.1.1
Testing
Test the pump as given in Section 5.