1.2
Technical Description
The
Flowtron
®
DVT
System has been specifically designed to provide
prophylactic therapy for patients at risk of developing Deep Vein Thrombosis.
For optimum results it is recommended that
Flowtron DVT
be applied
pre-operatively and used continuously during the pre-, intra-, and post-operative
periods.
The system comprises :
1.
Pressure regulator pump Model AC500/2
2.
Suitable patient garments from the
Flowtron DVT
range
3.
L550 Tubeset with snaplock connectors
Pressure regulated pump unit AC500/2 to be used with L550 and
Flowtron DVT
garments only.
The pump inflates and deflates the garments on a one-minute cycle
(approximately 12 seconds inflating, 48 seconds deflating). During the cycle,
air is pumped to one outlet followed by the commencement of inflation at the
second outlet approximately 30 seconds later.
The pump unit is piece of precision equipment, and provided that it is handled
and serviced correctly (normally every 12 months), should give long and trouble
free service.
Controls and indicators on the pump unit comprise:
a)
Power switch (green rocker switch) illuminated when in the ‘ON’
position
b)
Pressure control knob. Pressure adjustment between 20 mmHg
and 60 mmHg
c)
Red LED alarm indicator and audible signal indicates LOW and
HIGH pressure faults
Note:
The
Flowtron DVT
pump is fitted with both audible and visual alarm
indicators. When a fault occurs the audible alarm sounds and the
LED illuminates.
F048
1-2
Issue 3
Mar 2003
Содержание Flowtron DVT
Страница 1: ...F048 Issue 3 Mar 2003 Huntleigh HEALTHCARE Flowtron DVT Service Manual ...
Страница 2: ...F048 Issue 3 Mar 2003 ...
Страница 6: ...Issue 3 iv F048 Mar 2003 ...
Страница 10: ...F048 1 4 Issue 3 Mar 2003 ...
Страница 12: ...F048 2 2 Issue 3 Mar 2003 ...
Страница 16: ...F048 3 4 Issue 3 Mar 2003 ...
Страница 18: ...F048 4 2 Issue 3 Mar 2003 ...
Страница 30: ...F048 5 12 Issue 3 Mar 2003 ...
Страница 32: ...F048 6 2 Issue 3 Mar 2003 ...
Страница 33: ...WIRING DIAGRAMS F048 1 1 Issue 3 Mar 2003 Flowtron DVT AC 500 ...
Страница 34: ...F048 1 2 Issue 3 Mar 2003 ...
Страница 37: ...F048 Issue 3 Mar 2003 ...