Three-phase energy meter field calibrator
screen
Notes:
1) Before correcting the clip-on ammeter, the best clean the clamp mouth;
2) Keep the clip-on ammeter as far as possible from high current, otherwise it will affect the
accuracy of correction;
3) Correcting the clip-on ammeter lasts about 3 minutes, during which the instrument can not
be shut down, and other operations can not be carried out; when selecting the battery
power supply, ensure that the battery has at least one section of electric quantity;
4) When correcting the clip-on ammeter, please do not connect the photoelectric signal;
5) When correcting the clip-on ammeter, please do not connect the terminal current.