6.7.1. Description of anomalous results
1.
Once acquisition is
completed, if the instrument
has detected an uncorrect
phase sequence, it displays
a screen similar to the one
reported here to the side and
gives a long acoustic signal
1 2 3
132
Correct phase sequence
NOT OK
1T
Func
2.
The results displayed can be saved by pressing the
SAVE
key twice or the
SAVE
key and, subsequently, the
ENTER
key (§ 8.1)
3. If between the acquisition of
the first and second voltage
a time higher than the limit
time has elapsed, the
instrument displays a screen
similar to the one reported
here to the side and gives a
long acoustic signal
1 2 3
1--
Time out
1T
Func
4. If, during voltage acquisition,
the instrument detects an
input voltage higher than the
maximum limit value, it
displays a screen similar to
the one reported here to the
side and stops measuring
1 2 3
---
Vin > Vmax
2T
Func
5.
The previous anomalous results cannot be saved
6.8.
AUX:
REAL TIME MEASUREMENT OF THE ENVIRONMENTAL
PARAMETERS
By means of external transducers, this function allows measuring the following
environmental parameters: