![HT 400 Series Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/ht/400-series/400-series_user-manual_2178071029.webp)
400
Series
EN - 27
Fig. 12: Instrument connection for 400V + N + PE three-phase RCD test by means of
single cables and remote probe
Fig. 13: Instrument connection for 400V + N (no PE) three-phase RCD test by means of
single cables and remote probe
Fig. 14: Instrument connection for 400V + PE (no N) three-phase RCD test by means of
single cables and remote probe
1.
Press the
MENU
key, move the cursor to
RCD
in the main menu by means of the arrow keys
(
,
) and confirm with
ENTER
. Subsequently
the instrument displays a screen similar to the
one reported here to the side.
R C D
0°
---ms
FRQ=50.0Hz
Ut=0.0V
VP-N=230V
VP-Pe=230V
x1 30mA
50V
Func
IdN RCD UL
2.
Use the
,
keys to select the parameter to be modified, and the
,
keys
to modify the parameter value.
It is not necessary to confirm the selection with ENTER
.
Func
The virtual Func key allows setting the measuring mode of the
instrument, which may be:
AUTO, x½, x1, x2, x5, , RA
IdN
The virtual IdN key allows setting the nominal value of the RCD’s
tripping current, which may be:
10mA, 30mA, 100mA, 300mA,
500mA, 650mA, 1A