![HT 400 Series Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/ht/400-series/400-series_user-manual_2178071023.webp)
400
Series
EN - 21
5. If the instrument detects a
voltage value higher than
10V at the input leads, the
screen reported here to the
side is displayed
L O W
- - - -
R +
R -
- - - -
- - - -
- - - m A
- - - m A
Vin > Vlim
AUTO
4.00
0.21
Func
Lim CAL
6. In case it was detected that
the calibrated resistance is
higher than the measured
resistance increased by
0.05
(R
CAL
>R
MEAS
+0.05
),
the instrument gives a long
acoustic signal and displays
a screen similar to the one
reported here to the side
L O W
0 . 0 0
R +
R -
0 . 0 0
0 . 0 0
2 1 4 m A
2 1 4 m A
CAL > RES
AUTO
4.00
0.21
Func
Lim CAL
7. By using the CAL mode, if
the instrument detects a
resistance higher than 5
at
input leads, a screen similar
to the one reported here to
the side is displayed, and the
instrument remains in the
condition of no resistance
calibrated
L O W
- - - -
R +
R -
- - - -
- - - -
- - - m A
- - - m A
Reset value
CAL
4.00
0.21
Func
Lim CAL
8. By using the CAL mode,
verifying the calibrated value
at the end of the CAL
procedure, if the condition:
R
CAL
≤
R
MEAS
≤
R
CAL
+ 0.05
Ω
is not met, a screen similar
to the one reported here to
the side is displayed, and the
instrument remains in the
condition of no resistance
calibrated
L O W
1 . 9 8
R +
R -
1 . 9 8
1 . 9 8
2 1 0 m A
2 1 0 m A
Not correct
CAL
4.00
----
Func
Lim CAL
9.
The previous anomalous results cannot be saved