400
Series
EN - 15
3. If the measured phase-to-
earth insulation value is
lower than the set limit, the
instrument displays a screen
similar to the one reported
here to the side and gives a
long acoustic signal
A U T O
R a = 1 7 8 9
T r c d > 9 9 9 m s
R P - P e = 0 . 0 1 M
Value of earth resistance
Value of the RCD’s tripping time
Value of the phase-to-earth
insulation resistance
NOT OK
30mA
50V 500V
IdN
RCD UL VNom
4.
The results displayed can be saved by pressing the
SAVE
key twice or the
SAVE
key and, subsequently, the
ENTER
key (§ 8.1)
5. If the instrument detects that
the phase and neutral leads
are inverted, the message
reported here to the side is
displayed. Rotate the shuko
plug or check the connection
of the single cables
A U T O
R a = - - - -
T r c d = - - - - m s
R P - P e = - - - - M
REVERSE P-N
The phase and neutral conductors
are inverted
30mA
50V 500V
IdN
RCD UL VNom
6. If the instrument detects that
the phase and earth leads
are inverted, the message
reported here to the side is
displayed. check the
connection of the cables
A U T O
R a = - - - -
T r c d = - - - - m s
R P - P e = - - - - M
REVERSE P-PE
The phase and earth conductors are
inverted
30mA
50V 500V
IdN
RCD UL VNom
7. If the instrument detects a
phase-to-neutral voltage and
a phase-to-earth voltage
lower than the limit, the
message reported here to
the side is displayed. Check
that the system being tested
is energized
A U T O
R a = - - - -
T r c d = - - - - m s
R P - P e = - - - - M
Low voltage
Insufficient voltage
30mA
50V 500V
IdN
RCD UL VNom