IFU EN T800, Rev 09/2016-07-07, art.no. 8004007
20
Calibration
Perform a pressure test for internal leakage and a pressure calibration of the system each twelve
(12) months or after changing any vital parts in the unit such as internal hoses, pumps or valves.
Contact your local distributor for more information about leakage test.
Battery
The battery needs to be checked once a year. Check that a dis-charged battery can be charged
within 3 hours. If not, the battery needs to be replaced (the battery is considered as a
consumable). We recommend that the battery is replaced every twentyfour (24) months.
7 Data log
The following data are logged for each operation: Date and time of start, set pressure, actual
operation duration time, mains/IVRA/cuff status description and error description.
The log for the recent operation can be printed out only after a procedure.
Note:
The log is maintained during an AC power fail and after the Tourniquet is powered down.
8 Technical specifications
Electrical power input: 100-240 VAC, 50-60 Hz
AC receptacle: IEC type per CEE22V (hospital grade)
Temperature range: ope5 till+40 °C,
Temperature range: transport & storing -10 to +50 °C.
Relative Humidity range: 20-80% non-condensing (operating, transport, storing)
Atmospheric pressure range: 75 to 106 kPa
Pressure, maximal level: 600 mmHg (emergency stop during malfunction)
Pressure, operating range: 50 mmHg to 550 mmHg
Weight: 2.5 kg
Dimensions: 200 mm (wide) x 195 mm (high) x 145 mm (deep)
Pressure output port: CPC connector
Accuracy: ±5 mmHg or 2% full scale
Inflate time: ~5 seconds
Class & type: Class 1, type B
Printer interface: USB
Backup battery: NiMh, 7.2 V 4500 mAh
Alarm sound level: 73dB