IFU EN T800, Rev 09/2016-07-07, art.no. 8004007
18
Alarm message Description
Priority Comment
Timeout!
Timer counting down
Medium
Time > 60
Timer 60 minutes
Low
Time > 90
Timer additional 30
minutes
Low
Time > 90 n* 15
min
Timer reminder every 15
minutes
Low
Time limit
Timer after 180 minutes
Medium
Risk for harming patient
Fatal CPU failure
Instrument testing
continuously
High
Instrument error:
Restart instrument, or call service
Fatal pressure
failure
Instrument testing
continuously
High
Instrument error:
Restart instrument, or call service
Mains failure
Mains failure in Idle (cuff
has no pressure)
Low
Safe state until mains returns
Instrument error
Instrument testing
continuously
Medium
Instrument error:
Restart instrument, or call service
Mains failure
For 800-x0 model: Mains
failed and cuff has pressure Low
Clears when mains returns
Critical battery level Battery level < 25%
Low
Safe state:
Connect mains to charge battery
Critical battery level During start up
Low
Connect mains for 0-2 hours to charge
battery, until battery indicator shows full
battery capacity
Low battery level
Battery level < 50%
Low
Connect mains to charge battery
System pressure
failure (infl)
Instrument testing
continuously
Medium
Instrument error:
Restart instrument, or call service
Pump failure
(inflate)
Instrument testing
continuously
Medium
Instrument error:
Restart instrument, or call service
Leakage (inflate)
Leakage while inflating
Low
Check tubing and cuff
Kink (inflate)
Kink while inflating
Low
Safe state:
Acknowledge alarm or perform selftest
to return to normal operation
Critical pressure
Pressure level < 100 mmHg
below Pset if IVRA
High
Safe state:
Acknowledge alarm or perform selftest
to return to normal operation
Critical pressure
Pressure level < 100 mmHg
below Pset if not IVRA
Medium
Safe state:
Acknowledge alarm or perform selftest
to return to normal operation
Low pressure
Pressure level < 15 mmHg
below Pset
Low
Clears if pressure returns
High pressure
Pressure level > 30 mmHg
above Pset
Low
Clears if pressure returns
Critical high
pressure
Pressure level > 100 mmHg
above Pset
Medium
Safe state:
Acknowledge alarm or perform selftest
to return to normal operation
Leakage
Leakage while inflated
Medium
Battery fault
Battery cannot be charged
properly
High
Instrument error:
Restart instrument, or call service
Sensor consistency
Pressure sensor fail
detected
High
Instrument error: Restart instrument, or
call service
Cuff not connected A cuff is un-connected
Low
Connect cuff