9 Accessories
Printer:
Art. no 800-10 Model EPSON TM-T88V supplied with screened cable.
Clamp:
Art. no 800-25.
Mobile stand:
Art. no 400-45.
Basket:
Art. no 400-85.
Hpm™ Reusable and autoclavable single and double cuffs, Hpm™ Disposable single and double
cuffs. Hpm™ disposable cuffs are sterilized and are only to be used according to hospital
procedures and regulations. The following codes are referred to as
Hpm™ Disposable cuffs
;
Hpm™ single disposable cuffs:
15-100, 15-110, 15-120, 15-130, 15-140, 15-150, 15-160, 15-
170 & 15-180.
Hpm™ single disposable cuffs conical:
75-140, 75-150, 75-160 & 75-170.
Hpm™ double disposable cuffs:
25-100, 25-110, 25-120, 25-140 & 25-160.
Hpm™ Tourniquet 800-series are only to be used together with cuffs marked according to the
CE standard.
We recommend that you only use Hpm™ CE-marked cuffs with Hpm™ Tourniquet 800-20 and
800-40.
10 Appendix EMC guidance
The following guidance information are required by EN 60601-1-2 (2007):
Warning
The tourniquet should not be used adjacent to or stacked with other equipment.
Warning
The use of any other printer than listed in chapter 9 may result in increased emisssions or
decreased immunity of the tourniquet.
Essential performance (as defined in EN60601-1-2)
An alarm system should signal whenever the tourniquet inflated pressure is 30 mmHg
higher or 15 mmHg lower than the set pressure.
In single fault condition the cuffs should be isolated using a block valve with a maximum
leakage of 1 mmHg/min.
The tourniquet inflated pressure should not be affected by disturbances (excluding
external pressure changes) by more than +30 to -15 mmHg.
The tourniquet inflated pressure should compensate for pressure changes due to
external pressure changes on the cuff within a few seconds