IFU EN T800, Rev 09/2016-07-07, art.no. 8004007
8
The power cord plug must be removed to break AC current. The wall socket outlet
should be installed nearby the tourniquet and be easily accessible.
2.3
Procedure options
The Hpm™ Tourniquet has two intended options of use, single cuff procedure and double cuff
IVRA procedure. When a single cuff is used, the user can inflate or deflate the applied cuff for
either upper or lower limb. When IVRA procedures are performed, the proximal and distal sides
are synchronized regarding set pressure and time, in order to perform safe surgeries. The “IVRA”
icon is activated before the start of the surgery and will be indicated with a green illuminated
“IVRA” icon.
2.4
Unit start up
Make sure that the cuffs are disconnected.
The Tourniquet is turned on by pressing the ON/OFF button on the backside of the
Tourniquet.
The tourniquet will then perform a Self Test of system, pump function and a leakage test.
A sensor calibration is performed against the atmospheric pressure.
After this you will be asked to press the green OK icon.
If you tap outside of the green
OK icon, you will be asked to calibrate the screen. Conduct the calibration by following
the instructions on the screen.
After pressing the OK icon you will then hear two (2) beeps, one from the speaker and one from
the back-up alarm.
Press OK to confirm hearing these two (2) beeps. If the user answers “CANCEL” two (2) times
the tourniquet goes into an instrument error-mode (code 20) and needs to be restarted.
The tourniquet is now ready to use.
Main menu display when the tourniquet is ready for use.
The tourniquet is now ready for operation with the pressure (ref. 3.2) and time (ref. 3.1.2) earlier
set by the user. Cuff inflation is done by pressing the “Inflate” icon.
To perform a manual self test, please enter the sub menu by pressing
(sub menu icon) and
then press
(self test icon) (ref. 3.1.3).