IFU EN T800, Rev 09/2016-07-07, art.no. 8004007
22
Guidance and manufacturer's declaration - electromagnetic emissions
The tourniquet is intended for use in the electromagnetic environment specified below.
The user should assure that it is used in such an environment
Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The device uses RF energy only for its internal function.
Therefor, its RF emmissions ar very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The tourniquet is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonics
IEC61000-3-2
Not
Applicable
Voltage fluctuations
IEC 61000-3-3
Not
Applicable
Guidance and manufacturer's declaration - electromagnetic immunity
The tourniquet is intended for use in the electromagnetic environment specified below.
The user should assure that it is used in such an environment
Immunity test
IEC 601 test level
Compliance level
Electromagnetic immunity
guidance
ESD
IEC 61000-4-2
±2 kV, ±4 kV and ±6 kV
contact discharges
±2 kV, ±4 kV and ±8 kV
air discharges
±2 kV, ±4 kV and ±6 kV
contact discharges
±2 kV, ±4 kV and ±8 kV
air discharges
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
shall be at least 30%
Radiated fields
EN 61 000-4-3
3 V/m
10 V/m
Mains power quality should be of that
of a typical commercial or hospital
environment.
Burst
IEC 61000-4-4
±2 kV
±2 kV
Mains power quality should be of that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±2 kV
±2 kV
Mains power quality should be of that
of a typical commercial or hospital
environment.
Conducted
distrurbances
IEC 61000-4-6
3 V with 80% AM @ 1
kHz
3 V with 80% AM @ 1
kHz
Mains power quality should be of that
of a typical commercial or hospital
environment.
Power frequency
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristics of a
typical location in a typical commercial
or hospital environment.
Voltage dips and
interruptions
IEC 61000-4-11
Reduction:
> 95% for 10 ms
60% for 100 ms
30% for 500 ms
> 95% for 5 s
Reduction:
> 95% for 10 ms
60% for 100 ms
30% for 500 ms
> 95% for 5 s
Mains power quality should be of that
of a typical commercial or hospital
environment. If the user of the
Tourniquet requires continued
operations during power mains
interruptions, it is recommended that
the tourniquet be powered from an
uninterruptible power supply or
battery.
Содержание Tourniquet 800-20
Страница 1: ...Manual Hpm Tourniquet 800 20 800 40 English ...
Страница 2: ...IFU EN T800 Rev 09 2016 07 07 art no 8004007 ...
Страница 27: ...IFU EN T800 Rev 09 2016 07 07 art no 8004007 ...