![HP ESG-D Series Скачать руководство пользователя страница 314](http://html.mh-extra.com/html/hp/esg-d-series/esg-d-series_user-manual_160473314.webp)
5-178
Hardkey and Softkey Reference
Mode - GSM
Status after Normal Preset: 3A
SCPI Commands:
[:SOURce]:RADio:GSM:SLOT0|[1]|2|3|4|5|6|7:ACCess:ETail <bit_pattern>
[:SOURce]:RADio:GSM:SLOT0|[1]|2|3|4|5|6|7:ACCess:ETail?
Ext
Press this softkey in the data selection menus to select external data. With
Ext
selected, you should apply the data signal to the DATA INPUT connector. If you
have selected
Ext
as a data pattern for modulated transmissions that are not framed
(
Data Format Pattern Framed
is set to
Pattern
),
Ext
is displayed in the
Data
field
in the top line of the text area of the display. If you have selected
Ext
for
configuring the timeslot data fields,
EXT
is displayed in either the
Data
field or
the
E
field (encryption bits), both of which are located near the bottom of the text
area of the display. In framed mode, the external data is gated to the data regions
of the timeslot.
Another
Ext
softkey is located in the Frame Trigger and Pattern Trigger menus.
Press
Ext
in either of these two menus to select external triggering for your data
pattern. Apply a TTL input to the rear panel PATTERN TRIG IN connector to
trigger your frame or pattern data.
Softkey Location: Press
Mode
,
GSM
,
Data
,
Ext
or press
Mode
,
GSM
,
Data Format Pattern Framed
,
Configure Timeslot
,
Configure Normal
,
E
,
Ext
or press
Mode
,
GSM
,
Pattern Repeat Single Cont
,
Frame Trigger
,
Ext
or press
Mode
,
GSM
,
Data Format Pattern Framed
,
Frame Repeat Single Cont
,
Frame Trigger
,
Ext
Ext
is also located in the Configure Custom menu and in the E Menus found in the
Configure Sync menu and the Configure Access menu.
SCPI Commands:
[:SOURce]:RADio:GSM:DATA EXTernal
[:SOURce]:RADio:GSM:SLOT0|[1]|2|3|4|5|6|7:ACCess:ENCRyption EXTernal
[:SOURce]:RADio:GSM:SLOT0|[1]|2|3|4|5|6|7:CUSTom EXTernal
[:SOURce]:RADio:GSM:SLOT0|[1]|2|3|4|5|6|7:NORMal:ENCRyption EXTernal
[:SOURce]:RADio:GSM:SLOT0|[1]|2|3|4|5|6|7:SYNC:ENCRyption EXTernal
[:SOURce]:RADio:GSM:TRIGger[:SOURce] EXTernal
Содержание ESG-D Series
Страница 4: ...ii ...
Страница 34: ...Contents 30 ...
Страница 46: ...1 12 Preparing for Use Verifying Signal Generator Operation ...
Страница 138: ...5 2 Hardkey and Softkey Reference ...
Страница 164: ...5 28 Hardkey and Softkey Reference Amplitude ...
Страница 166: ...5 30 Hardkey and Softkey Reference Arrow Keys ...
Страница 168: ...5 32 Hardkey and Softkey Reference Display Contrast Keys ...
Страница 188: ...5 52 Hardkey and Softkey Reference FM ...
Страница 196: ...5 60 Hardkey and Softkey Reference Frequency ...
Страница 198: ...5 62 Hardkey and Softkey Reference Help ...
Страница 200: ...5 64 Hardkey and Softkey Reference Hold ...
Страница 202: ...5 66 Hardkey and Softkey Reference Incr Set ...
Страница 236: ...5 100 Hardkey and Softkey Reference LF Out ...
Страница 238: ...5 102 Hardkey and Softkey Reference Local ...
Страница 240: ...5 104 Hardkey and Softkey Reference Mod On Off ...
Страница 242: ...5 106 Hardkey and Softkey Reference Mode ...
Страница 336: ...5 200 Hardkey and Softkey Reference Mode GSM ...
Страница 422: ...5 286 Hardkey and Softkey Reference Mode PDC ...
Страница 466: ...5 330 Hardkey and Softkey Reference Mode PHS ...
Страница 532: ...5 396 Hardkey and Softkey Reference Numeric Keypad ...
Страница 554: ...5 418 Hardkey and Softkey Reference Power Switch ...
Страница 564: ...5 428 Hardkey and Softkey Reference Preset ...
Страница 570: ...5 434 Hardkey and Softkey Reference Pulse ...
Страница 574: ...5 438 Hardkey and Softkey Reference Recall ...
Страница 576: ...5 440 Hardkey and Softkey Reference Return ...
Страница 578: ...5 442 Hardkey and Softkey Reference RF On Off ...
Страница 584: ...5 448 Hardkey and Softkey Reference Save ...
Страница 604: ...5 468 Hardkey and Softkey Reference Sweep List ...
Страница 666: ...7 20 Operation Userfile Applications ...
Страница 675: ...8 9 Safety and Regulatory Declaration of Conformity Declaration of Conformity ...