3-53
Troubleshooting
ESG-D Series Signal Generator Instrument-Specific Error Messages
Configuration error; Invalid option board
configuration.
This indicates that an invalid combination of option boards has
been configured. If this error has occurred as the result of a
customer-installed option, uninstall all options and then
reinstall the correct options. If the error persists, contact the
factory.
700
State Save Recall Error;Recall aborted. Unable
to recall the state from non-volatile memory.
This indicates that the state file was not readable, so the recall
was aborted. If state file exists, delete explicitly using the
memory catalog.
State Save Recall Error;Recalled state has a
bad checksum. No instrument state change.
This indicates that the state file was corrupt or out-of-date, so
the recall was ignored. If state file exists, delete explicitly using
the memory catalog.
State Save Recall Error;Recall data different
from FW revision. No instrument state change.
Indicates that an attempt was made to recall a state that was
saved with an incompatible version of the instrument firmware.
This typically occurs when a state file is copied from an
instrument with a newer version of firmware to an instrument
with an older version of firmware. Newer versions of
instrument firmware can read older state files.
State Save Recall Error;Recall non-saved state
register. Recall ignored.
Indicates that a recall was attempted for a state register that is
unused. If state file exists, delete explicitly using catalog.
State Save Recall Error;Delete sequence
<sequence_name> ignored.
Indicates that a
STATE:
file in a sequence that is being deleted
could not be deleted due to the file not being found, data
corruption, etc. If state file exists, delete explicitly using the
memory catalog.
Содержание ESG-D Series
Страница 4: ...ii ...
Страница 34: ...Contents 30 ...
Страница 46: ...1 12 Preparing for Use Verifying Signal Generator Operation ...
Страница 138: ...5 2 Hardkey and Softkey Reference ...
Страница 164: ...5 28 Hardkey and Softkey Reference Amplitude ...
Страница 166: ...5 30 Hardkey and Softkey Reference Arrow Keys ...
Страница 168: ...5 32 Hardkey and Softkey Reference Display Contrast Keys ...
Страница 188: ...5 52 Hardkey and Softkey Reference FM ...
Страница 196: ...5 60 Hardkey and Softkey Reference Frequency ...
Страница 198: ...5 62 Hardkey and Softkey Reference Help ...
Страница 200: ...5 64 Hardkey and Softkey Reference Hold ...
Страница 202: ...5 66 Hardkey and Softkey Reference Incr Set ...
Страница 236: ...5 100 Hardkey and Softkey Reference LF Out ...
Страница 238: ...5 102 Hardkey and Softkey Reference Local ...
Страница 240: ...5 104 Hardkey and Softkey Reference Mod On Off ...
Страница 242: ...5 106 Hardkey and Softkey Reference Mode ...
Страница 336: ...5 200 Hardkey and Softkey Reference Mode GSM ...
Страница 422: ...5 286 Hardkey and Softkey Reference Mode PDC ...
Страница 466: ...5 330 Hardkey and Softkey Reference Mode PHS ...
Страница 532: ...5 396 Hardkey and Softkey Reference Numeric Keypad ...
Страница 554: ...5 418 Hardkey and Softkey Reference Power Switch ...
Страница 564: ...5 428 Hardkey and Softkey Reference Preset ...
Страница 570: ...5 434 Hardkey and Softkey Reference Pulse ...
Страница 574: ...5 438 Hardkey and Softkey Reference Recall ...
Страница 576: ...5 440 Hardkey and Softkey Reference Return ...
Страница 578: ...5 442 Hardkey and Softkey Reference RF On Off ...
Страница 584: ...5 448 Hardkey and Softkey Reference Save ...
Страница 604: ...5 468 Hardkey and Softkey Reference Sweep List ...
Страница 666: ...7 20 Operation Userfile Applications ...
Страница 675: ...8 9 Safety and Regulatory Declaration of Conformity Declaration of Conformity ...