5-78
Hardkey and Softkey Reference
I/Q
Q Offset
Press this softkey to enter an origin offset voltage for internally-generated or
externally-applied quadrature-phase signals. Enter the value as a percentage of
full scale (500 mV). The range of values allowed is
−
100% to +100%. Change the
origin offset to adjust out imperfections in the quadrature-phase signal or to
introduce calibrated impairments. Set the origin offset as desired and then toggle
the
I/Q Adjustments Off On
softkey to
On
.
Softkey Location: Press
I/Q
,
I/Q Adjustments
,
Q Offset
Status after Normal Preset: 0.0%
SCPI Commands:
[:SOURce]:DM:IQADjustment:QOFFset <val><unit>
[:SOURce]:DM:IQADjustment:QOFFset?
Revert to Default Cal Settings
Press this softkey to restore the original factory calibration data for the internal
I/Q modulator. The latest calibration data from the user calibration is discarded.
Although the factory calibration data may not be optimized for the signal
generator’s present status, the data can be used as a starting point if the current
data is lost (if the I/Q calibration is unable to execute).
Softkey Location: Press
I/Q
,
More (1 of 2)
,
I/Q Calibration
,
Revert to Default Cal Settings
SCPI Commands:
:CALibration:IQ:DEFault
Содержание ESG-D Series
Страница 4: ...ii ...
Страница 34: ...Contents 30 ...
Страница 46: ...1 12 Preparing for Use Verifying Signal Generator Operation ...
Страница 138: ...5 2 Hardkey and Softkey Reference ...
Страница 164: ...5 28 Hardkey and Softkey Reference Amplitude ...
Страница 166: ...5 30 Hardkey and Softkey Reference Arrow Keys ...
Страница 168: ...5 32 Hardkey and Softkey Reference Display Contrast Keys ...
Страница 188: ...5 52 Hardkey and Softkey Reference FM ...
Страница 196: ...5 60 Hardkey and Softkey Reference Frequency ...
Страница 198: ...5 62 Hardkey and Softkey Reference Help ...
Страница 200: ...5 64 Hardkey and Softkey Reference Hold ...
Страница 202: ...5 66 Hardkey and Softkey Reference Incr Set ...
Страница 236: ...5 100 Hardkey and Softkey Reference LF Out ...
Страница 238: ...5 102 Hardkey and Softkey Reference Local ...
Страница 240: ...5 104 Hardkey and Softkey Reference Mod On Off ...
Страница 242: ...5 106 Hardkey and Softkey Reference Mode ...
Страница 336: ...5 200 Hardkey and Softkey Reference Mode GSM ...
Страница 422: ...5 286 Hardkey and Softkey Reference Mode PDC ...
Страница 466: ...5 330 Hardkey and Softkey Reference Mode PHS ...
Страница 532: ...5 396 Hardkey and Softkey Reference Numeric Keypad ...
Страница 554: ...5 418 Hardkey and Softkey Reference Power Switch ...
Страница 564: ...5 428 Hardkey and Softkey Reference Preset ...
Страница 570: ...5 434 Hardkey and Softkey Reference Pulse ...
Страница 574: ...5 438 Hardkey and Softkey Reference Recall ...
Страница 576: ...5 440 Hardkey and Softkey Reference Return ...
Страница 578: ...5 442 Hardkey and Softkey Reference RF On Off ...
Страница 584: ...5 448 Hardkey and Softkey Reference Save ...
Страница 604: ...5 468 Hardkey and Softkey Reference Sweep List ...
Страница 666: ...7 20 Operation Userfile Applications ...
Страница 675: ...8 9 Safety and Regulatory Declaration of Conformity Declaration of Conformity ...