Comments
•
The last frequency point is output for the same length of time as all
other points. The
SWEep:TIME
value is the time from the start of
each sweep until the last frequency begins to be output and does not
include the time for the last frequency point. Therefore, if a specific
sweep repetition time is desired,
SWEep:TIME
should be set
according to the following equation:
SWEep:TIME =
time
* (
points
- 1) /
points
Thus, to set a repetition time of 1 S for a 100 point sweep,
SWEep:TIME should be set to .99 S.
•
Executable when initiated: Query form only
•
Coupled command: Yes
•
Related Commands:
SOURce:SWEep:POINts, CAL:STATe:AC
•
*RST Condition:
SOURce:SWEep:TIME 1.05
Example
Setting the duration of the sweep
SWE:TIME 20
Set sweep to take 20 seconds
[SOURce:]SWEep:TIME
176 HP E1340A Command Reference
Chapter 7
Содержание E1340A
Страница 12: ...Notes 12 HP E1340A Arbitrary Function Generator Module User s Manual ...
Страница 14: ...14 HP E1340A Arbitrary Function Generator Module User s Manual ...
Страница 42: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 42 ...
Страница 54: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 54 ...
Страница 57: ...Figure 3 1 Generating Arbitrary Waveforms 57 Generating Arbitrary Waveforms with the HP E1340A Chapter 3 ...
Страница 58: ...Chapter 3 Generating Arbitrary Waveforms with the HP E1340A 58 ...
Страница 84: ...Chapter 4 HP E1340A Sweeping and Frequency Shift Keying 84 ...
Страница 130: ...Chapter 6 HP E1340A High Speed Operation 130 ...
Страница 202: ...202 HP E1340A SCPI Conformance Information Chapter 7 ...
Страница 204: ...Figure 8 1 HP E1340A Status Groups and Associated Registers Chapter 8 HP E1340A AFG Status 204 ...
Страница 218: ...218 HP E1340A Specifications Appendix A ...
Страница 284: ...284 HP E1340A Register Based Programming Appendix C ...
Страница 295: ...Index HP E1340A Arbitrary Function Generator User s Manual 295 ...