Code
Message
Description
-161
Invalid block data
The number of bytes in a definite length data block
does not equal the number of bytes indicated by the
block header.
-168
Block data not allowed
Block data was specified when another parameter
type (i.e. discrete, numeric, boolean) is required.
-178
Expression data not allowed
The parameter was specified as an expression (e.g.
SOUR:FREQ1:FIX (A*B).
-183
Invalid inside macro definition
Voltage or segment list is inside a macro.
-213
Init ignored
INITiate:IMMediate received while the AFG was
currently initiated.
-221
Settings conflict
See "Settings Conflict Error Messages" at the end of
this table.
-222
Data out of range
Parameter value is out of range for any AFG
configuration (e.g. SOUR:FREQ1:FIX 1E9).
-240
Hardware error
The HP E1340A is not responding. Possible
hardware failure.
-270
Macro error
*RMC <name> was executed and name is not
defined.
-272
Macro execution error
Macro program data sequence could not be executed
due to a syntax error within the macro definition.
-273
Illegal macro label
The macro label defined in the *DMC command was
too long, the same as a common command keyword,
or contained invalid header syntax.
-276
Macro recursion error
A macro program data sequence could not be
executed because the sequence leads to the
execution of a macro being defined.
-277
Macro redefinition not allowed
A macro label in the *DMC command could not be
executed because the macro label was already
defined.
-330
Self-test failed
Note the information associated with the message for
a description of the failure.
-350
Too many errors
The HP E1340A A error queue is full and additional
errors have occurred.
-410
Query INTERRUPTED
The HP E1340A was sent a command before it was
finished responding to a query command.
-420
Query UNTERMINATED
The controller (computer) attempts to read a query
response from the HP E1340A without having first
sent a complete query command.
Table B-4. HP E1340A Error Messages (Cont’d)
Appendix B
HP E1340A Useful Tables 225
Содержание E1340A
Страница 12: ...Notes 12 HP E1340A Arbitrary Function Generator Module User s Manual ...
Страница 14: ...14 HP E1340A Arbitrary Function Generator Module User s Manual ...
Страница 42: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 42 ...
Страница 54: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 54 ...
Страница 57: ...Figure 3 1 Generating Arbitrary Waveforms 57 Generating Arbitrary Waveforms with the HP E1340A Chapter 3 ...
Страница 58: ...Chapter 3 Generating Arbitrary Waveforms with the HP E1340A 58 ...
Страница 84: ...Chapter 4 HP E1340A Sweeping and Frequency Shift Keying 84 ...
Страница 130: ...Chapter 6 HP E1340A High Speed Operation 130 ...
Страница 202: ...202 HP E1340A SCPI Conformance Information Chapter 7 ...
Страница 204: ...Figure 8 1 HP E1340A Status Groups and Associated Registers Chapter 8 HP E1340A AFG Status 204 ...
Страница 218: ...218 HP E1340A Specifications Appendix A ...
Страница 284: ...284 HP E1340A Register Based Programming Appendix C ...
Страница 295: ...Index HP E1340A Arbitrary Function Generator User s Manual 295 ...