STATus
The
STATus
subsystem controls the SCPI-defined Operation and
Questionable Signal Status Registers. Each is comprised of a Condition
Register, an Event Register, an enable mask, and negative and positive
transition filters.
Each Status Register works as follows: when a condition occurs, the
appropriate bit in the Condition Register is set or cleared. If the the
corresponding transition filter is enabled for that bit, the same bit is set in
the associated Event Register. The contents of the Event Register and the
enable mask are logically ANDed bit-for-bit; if any bit of the result is set,
the summary bit for that register is set in the status byte. The status byte
summary bit for the Operation Status Register is bit 7; for the Questionable
Signal Status Register, bit 3.
Operation Status
Register
Only bits 0 (
CALibrating
), 3 (
SWEeping
), and 8 (
INITiated
) are defined for
the HP E1340A. All other bits are always zero.
Bit 0 - CALibrating:
Set (1) during calibration. Cleared (0)
otherwise.
Bit 3 - SWEeping:
Set (1) while a frequency sweep or list is in
progress. Cleared (0) when waveform generation is halted, when
frequency sweeping or lists are not selected, and at the end of each
sweep or list.
Bit 8 - INITiated:
Set (1) by the
INITiate:IMMediate
command.
Cleared (0) by the
ABORt
command. When outputting a single
burst of waveform repetitions (
ARM:LAYer2:COUNt 1
set and
ARM:LAYer1:COUNt
not set to
INFinity
), it is also cleared by any
other command executed after waveform generation completes.
Questionable Signal
Status Register
All bits are always 0. This register is implemented only for SCPI
compatibility purposes.
Subsystem Syntax
STATus
:OPERation|QUEStionable
:CONDition?
[query only]
:ENABle <
unmask
>
[:EVENt]?
[query only]
:NTRansition <
unmask
>
:PTRansition <
unmask
>
:PRESet
[no query]
STATus[:LEVel][:IMMediate]:OFFSet
Chapter 7
HP E1340A Command Reference 181
Содержание E1340A
Страница 12: ...Notes 12 HP E1340A Arbitrary Function Generator Module User s Manual ...
Страница 14: ...14 HP E1340A Arbitrary Function Generator Module User s Manual ...
Страница 42: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 42 ...
Страница 54: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 54 ...
Страница 57: ...Figure 3 1 Generating Arbitrary Waveforms 57 Generating Arbitrary Waveforms with the HP E1340A Chapter 3 ...
Страница 58: ...Chapter 3 Generating Arbitrary Waveforms with the HP E1340A 58 ...
Страница 84: ...Chapter 4 HP E1340A Sweeping and Frequency Shift Keying 84 ...
Страница 130: ...Chapter 6 HP E1340A High Speed Operation 130 ...
Страница 202: ...202 HP E1340A SCPI Conformance Information Chapter 7 ...
Страница 204: ...Figure 8 1 HP E1340A Status Groups and Associated Registers Chapter 8 HP E1340A AFG Status 204 ...
Страница 218: ...218 HP E1340A Specifications Appendix A ...