8-19
Preset State and Memory Allocation
Memory Allocation
If correction is on at the time of an external store, the calibration set is stored to disk.
(Note that inactive calibrations are not stored to disk.) When an instrument state is loaded
into the analyzer from disk, the stimulus and response parameters are restored first. If
correction is on for the loaded state, the analyzer will load a calibration set from disk that
carries the same title as the one stored for the instrument state.
Conserving Memory
If you are concerned about conserving memory, either internal memory or external disk
space, some of the most memory-intensive operations include:
• two-port error correction
• interpolated error correction
• 1601 measurement points
• using time domain
• saving data arrays and graphics with the instrument state
Using Saved Calibration Sets
When you are saving to internal memory (CMOS, non-volatile memory), calibration sets
are linked to the instrument state and measurement parameter for which the calibration
was done. Therefore a saved calibration can be used for multiple instrument states as long
as the measurement parameter, frequency range, and number of points are the same. A
full 2-port calibration is valid for any S-parameter measurement with the same frequency
range and number of points. When an instrument state is deleted from memory, the
associated calibration set is also deleted if it is unused by any other state.
The following hints will help you avoid potential problems:
• If a measurement is saved with calibration and interpolated calibration on, it will be
restored with interpolated calibration on.
• A calibration stored from one instrument and recalled by a different one will be invalid.
To ensure maximum accuracy, always recalibrate in these circumstances.
• No record is kept in memory of the temperature when a calibration set was stored.
Instrument characteristics change as a function of temperature, and a calibration
stored at one temperature may be inaccurate if recalled and used at a different
temperature. Refer to
Chapter 1 , “HP 8719/20/22ES Specifications and
and
Chapter 2 , “HP 8719/20/22ET Specifications and Characteristics,”
for allowable temperature ranges for individual specifications.
Содержание 8719ES
Страница 6: ...vi ...
Страница 10: ...Contents x Contents ...
Страница 11: ...1 1 1 HP 8719 20 22ES Specifications and Characteristics ...
Страница 60: ...1 50 HP 8719 20 22ES Specifications and Characteristics Instrument Specifications ...
Страница 61: ...2 1 2 HP 8719 20 22ET Specifications and Characteristics ...
Страница 98: ...2 38 HP 8719 20 22ET Specifications and Characteristics Instrument Specifications ...
Страница 99: ...3 1 3 Front Rear Panel ...
Страница 111: ...4 1 4 Menu Maps ...
Страница 113: ...4 3 Menu Maps Menu Maps Figure 4 2 Menu Map for Copy ...
Страница 114: ...4 4 Menu Maps Menu Maps Figure 4 3 Menu Map for Display ...
Страница 115: ...4 5 Menu Maps Menu Maps Figure 4 4 Menu Map for Format Figure 4 5 Menu Map for Local ...
Страница 116: ...4 6 Menu Maps Menu Maps Figure 4 6 Menu Map for Marker Marker Fctn and Marker Search ...
Страница 118: ...4 8 Menu Maps Menu Maps Figure 4 9 Menu Map for Power and Sweep Setup ET only ...
Страница 119: ...4 9 Menu Maps Menu Maps Figure 4 10 Menu Map for Power and Sweep Setup ES only ...
Страница 120: ...4 10 Menu Maps Menu Maps Figure 4 11 Menu Map for Preset ...
Страница 121: ...4 11 Menu Maps Menu Maps Figure 4 12 Menu Map for Save Recall ...
Страница 122: ...4 12 Menu Maps Menu Maps Figure 4 13 Menu Map for Scale Ref ...
Страница 128: ...5 1 5 Hardkey Softkey Reference ...
Страница 194: ...6 1 6 Error Messages ...
Страница 222: ...7 1 7 Options and Accessories ...
Страница 234: ...8 1 8 Preset State and Memory Allocation ...
Страница 253: ...8 20 Preset State and Memory Allocation Memory Allocation ...
Страница 254: ...9 1 9 Understanding the CITIfile Data Format ...
Страница 269: ...9 16 Understanding the CITIfile Data Format Useful Calculations ...
Страница 270: ...10 1 10 Determining System Measurement Uncertainties ...
Страница 281: ...10 12 Determining System Measurement Uncertainties Measurement Uncertainty Equations ...