6-6
Error Messages
Error Messages in Alphabetical Order
CANNOT READ/WRITE HFS FILE SYSTEM
Error Number
203
The disk is being accessed by the analyzer and is found to contain an HFS
(hierarchical file system) or files nested within subdirectories. The analyzer does
not support HFS. Replace the disk medium with a LIF or DOS formatted disk
that does not contain files nested within subdirectories.
CAN'T STORE/LOAD SEQUENCE, INSUFFICIENT MEMORY
Error Number
127
Your sequence transfer to or from a disk could not be completed due to
insufficient memory.
CAUTION: AUX CHANNELS MEASURE S-PARAMETERS ONLY
Error Number
216
This message appears if you try to assign a non-S-parameter measurement to
an auxiliary channel and a full 2-port cal is performed on channels 1 and 2 with
both channels uncoupled.
CAUTION: CORRECTION OFF: AUX CHANNEL(S) DISABLED
Error Number
215
This message is displayed when correction is forced off due to a stimulus change
that is not compatible with the current calibration while an auxiliary channel is
enabled. The auxiliary channels are restored when correction is turned on by
pressing
.
CH1 (CH2, CH3, CH4) TARGET VALUE NOT FOUND
Error Number
159
Your target value for the marker search function does not exist on the current
data trace.
CONTINUOUS SWITCHING NOT ALLOWED
Error Number
10
Your current measurement requires different power ranges on channel 1 and
channel 2. To protect the attenuator from undue mechanical wear, test set hold
will be activated. The "tsH" (test set hold) indicator in the left margin of the
display indicates that the inactive channel has been put in the sweep hold mode.
This message is also displayed if a mechanical switch test set is in use and
channels are measuring parameters that require the test set to switch
continuously, for example S11 on Channel 1 and S22 on Channel 2
COPY: device not responding; copy aborted
Error Number
170
The printer or plotter is not accepting data. Verify the cable connections, HP-IB
addresses, and otherwise ensure that the copy device is ready.
COPY OUTPUT COMPLETED
Information
Message
The analyzer has completed outputting data to the printer or plotter. The
analyzer can now accept another copy command.
Cal
CORRECTION on OFF
Содержание 8719ES
Страница 6: ...vi ...
Страница 10: ...Contents x Contents ...
Страница 11: ...1 1 1 HP 8719 20 22ES Specifications and Characteristics ...
Страница 60: ...1 50 HP 8719 20 22ES Specifications and Characteristics Instrument Specifications ...
Страница 61: ...2 1 2 HP 8719 20 22ET Specifications and Characteristics ...
Страница 98: ...2 38 HP 8719 20 22ET Specifications and Characteristics Instrument Specifications ...
Страница 99: ...3 1 3 Front Rear Panel ...
Страница 111: ...4 1 4 Menu Maps ...
Страница 113: ...4 3 Menu Maps Menu Maps Figure 4 2 Menu Map for Copy ...
Страница 114: ...4 4 Menu Maps Menu Maps Figure 4 3 Menu Map for Display ...
Страница 115: ...4 5 Menu Maps Menu Maps Figure 4 4 Menu Map for Format Figure 4 5 Menu Map for Local ...
Страница 116: ...4 6 Menu Maps Menu Maps Figure 4 6 Menu Map for Marker Marker Fctn and Marker Search ...
Страница 118: ...4 8 Menu Maps Menu Maps Figure 4 9 Menu Map for Power and Sweep Setup ET only ...
Страница 119: ...4 9 Menu Maps Menu Maps Figure 4 10 Menu Map for Power and Sweep Setup ES only ...
Страница 120: ...4 10 Menu Maps Menu Maps Figure 4 11 Menu Map for Preset ...
Страница 121: ...4 11 Menu Maps Menu Maps Figure 4 12 Menu Map for Save Recall ...
Страница 122: ...4 12 Menu Maps Menu Maps Figure 4 13 Menu Map for Scale Ref ...
Страница 128: ...5 1 5 Hardkey Softkey Reference ...
Страница 194: ...6 1 6 Error Messages ...
Страница 222: ...7 1 7 Options and Accessories ...
Страница 234: ...8 1 8 Preset State and Memory Allocation ...
Страница 253: ...8 20 Preset State and Memory Allocation Memory Allocation ...
Страница 254: ...9 1 9 Understanding the CITIfile Data Format ...
Страница 269: ...9 16 Understanding the CITIfile Data Format Useful Calculations ...
Страница 270: ...10 1 10 Determining System Measurement Uncertainties ...
Страница 281: ...10 12 Determining System Measurement Uncertainties Measurement Uncertainty Equations ...