![HP 8719ES Скачать руководство пользователя страница 214](http://html.mh-extra.com/html/hp/8719es/8719es_reference-manual_161529214.webp)
6-21
Error Messages
Error Messages in Alphabetical Order
SYNTAX ERROR
Error Number
33
You have improperly formatted an HP-IB command. Refer to the programmer’s
guide for proper command syntax.
SYST CTRL OR PASS CTRL IN LOCAL MENU
Error Number
36
The analyzer is in talker/listener mode. In this mode, the analyzer cannot
control a peripheral device on the bus. Use the local menu to change to system
controller or pass control mode.
TEST ABORTED
Error Number
113
You have prematurely stopped a service test.
TEST PORT OVERLOAD, REDUCE POWER
Error
Number 57
You have exceeded approxi14 dBm at one of the test ports (or
0 dBm at the A or B sampler, Option 012 only). When this occurs, reduce the
power to a lower level.
THIS LIST FREQ INVALID
Error
Number 133
You have set frequencies in the list that are outside of the allowable frequency
range of the analyzer. Reduce the frequency range of the list.
TOO MANY NESTED SEQUENCES. SEQ ABORTED
Error Number
164
You can only nest sequences to a maximum level of six. The sequence will abort
if you nest more than six.
TOO MANY SEGMENTS OR POINTS
Error Number
50
You can have a maximum of 30 segments or 1601 points in frequency list mode.
In power meter calibrations, you can have a maximum of 12 segments for power
sensor cal factors and power loss functions.
TRANSFORM, GATE NOT ALLOWED
Error Number
16
You can perform a time domain transformation only in linear and CW sweep
types.
TROUBLE! CHECK SETUP AND START OVER
Service Error
Number 115
Your equipment setup for the adjustment procedure in progress is not correct.
Check the setup diagram and instructions in the service guide. Start the
procedure again.
Содержание 8719ES
Страница 6: ...vi ...
Страница 10: ...Contents x Contents ...
Страница 11: ...1 1 1 HP 8719 20 22ES Specifications and Characteristics ...
Страница 60: ...1 50 HP 8719 20 22ES Specifications and Characteristics Instrument Specifications ...
Страница 61: ...2 1 2 HP 8719 20 22ET Specifications and Characteristics ...
Страница 98: ...2 38 HP 8719 20 22ET Specifications and Characteristics Instrument Specifications ...
Страница 99: ...3 1 3 Front Rear Panel ...
Страница 111: ...4 1 4 Menu Maps ...
Страница 113: ...4 3 Menu Maps Menu Maps Figure 4 2 Menu Map for Copy ...
Страница 114: ...4 4 Menu Maps Menu Maps Figure 4 3 Menu Map for Display ...
Страница 115: ...4 5 Menu Maps Menu Maps Figure 4 4 Menu Map for Format Figure 4 5 Menu Map for Local ...
Страница 116: ...4 6 Menu Maps Menu Maps Figure 4 6 Menu Map for Marker Marker Fctn and Marker Search ...
Страница 118: ...4 8 Menu Maps Menu Maps Figure 4 9 Menu Map for Power and Sweep Setup ET only ...
Страница 119: ...4 9 Menu Maps Menu Maps Figure 4 10 Menu Map for Power and Sweep Setup ES only ...
Страница 120: ...4 10 Menu Maps Menu Maps Figure 4 11 Menu Map for Preset ...
Страница 121: ...4 11 Menu Maps Menu Maps Figure 4 12 Menu Map for Save Recall ...
Страница 122: ...4 12 Menu Maps Menu Maps Figure 4 13 Menu Map for Scale Ref ...
Страница 128: ...5 1 5 Hardkey Softkey Reference ...
Страница 194: ...6 1 6 Error Messages ...
Страница 222: ...7 1 7 Options and Accessories ...
Страница 234: ...8 1 8 Preset State and Memory Allocation ...
Страница 253: ...8 20 Preset State and Memory Allocation Memory Allocation ...
Страница 254: ...9 1 9 Understanding the CITIfile Data Format ...
Страница 269: ...9 16 Understanding the CITIfile Data Format Useful Calculations ...
Страница 270: ...10 1 10 Determining System Measurement Uncertainties ...
Страница 281: ...10 12 Determining System Measurement Uncertainties Measurement Uncertainty Equations ...