4-8
HP 8511A Test Set Operating and Service Manual
Operation
Installation
Installation
Set each test set rear panel address switch to the address listed in Table 4-2
on page 4-7 if using a two test set configuration, and Table 4-3 on page 4-10
if configuring more than two test sets. Use the supplied test set interconnect
cable to connect test set number 1, J11 to the HP 8510. Use the supplied test
set interconnect cable to connect test set number 2, J11, to test set number 1,
J10. You may continue this test set “daisy chain” to include up to four test
sets if the total length of all test set interconnect cables does not exceed 13
meters (about 40 feet). The last test set in the chain does not require option
001.
If the RF coaxial switch(s) is not incorporated into the system, then the RF
input to the test set must be manually switched to the active test set.