HP 8511A Test Set Operating and Service Manual
2-3
Specifications
Dynamic Accuracy
Figure 2-2 and Figure 2-3 illustrate worst case magnitude and phase
uncertainty due to IF residuals and detector inaccuracies. This data excludes
uncertainty due to noise, frequency response, directivity, port matches and
connector repeatability.
Figure 2-2
Worst Case Dynamic Accuracy (Magnitude)
Figure 2-3
Worst Case Dynamic Accuracy (Phase)
*
Table 2-3
HP 8510/HP 8511A Characteristics
1
(part 2)
Tracking Drift (typical)
Magnitude 0.001 x V
°
C, Linear
Phase (0.1
+
f(GHz)) x V
°
C, degrees
Input Ports
Connector type:
Impedance:
Damage level:
Port input power for phase lock:
female 3.5 mm
50 ohms nominal
+
13 dBm (20 mW) CW RF input
2
Frequency:
0.045 to 8 GHz
8 to 20 GHz
20 to 26.5 GHz
Minimum:
−
40 dBm
−
38 dBm
−
35 dBm
Maximum:
−
5 dBm
−
5 dBm
−
5 dBm
1. The performance parameters listed are characteristic of the HP 8511A/HP 8510. They are typical or nominal figures and are not field verifiable.
2. Do not exceed
−
5 dBm input to sampler for proper phase lock operation.
*
Phase detector accuracy is better than 0.02 degrees, useful for measurements where only phase changes.